이미지 없음
이미지 없음
요청을 처리 중입니다.
Micro Motion-K-H-Series Hygienic Coriolis Meter H400S and 1500

Micro Motion H-Series Hygienic Coriolis Meter Consisting of H400S Sensor and 1500 Transmitter

사양

Micro Motion H400S Hygienic Coriolis Meter, 4 Inch (DN100), 316L Stainless Steel, 32 RA Finish
Mass flow accuracy (Liquid)
Up to ±0.05% of rate
Mass flow repeatability (Liquid)
Up to ±0.025% of rate
Volume flow accuracy (Liquid)
Up to ±0.05% of rate
Volume flow repeatability (Liquid)
Up to ±0.025% of rate
Mass flow accuracy (Gas)
Up to ±0.35% of rate
Mass flow repeatability (Gas)
Up to ±0.175% of rate
Density accuracy (Liquid)
Up to ±0.0005 g/cm³ (±0.5 kg/m³)
Temperature accuracy
±1 °C ±0.5% of reading
Sensor maximum working pressure
1,450 psig (100 barg)
Micro Motion 1500 Compact Control Room Single Variable Flow Transmitter
이용 가능 출력
모델 1500 : 유량 변수 1개가 필요한 응용 분야(mA 및 주파수/펄스 출력
모델 2500 : 출력과 더불어 여러 유량 변수를 동시에 모니터링해야 하는 응용 분야(mA, 주파수, Discrete I/O)
하우징
Polyamide PA 6.6
전력

•최소 19.2 - 28.8VDC
•6.3W
•설치(과전압) 카테고리 II, 오염 등급 2 요구 사항 충족
입출력 신호

•하나의 활성 4–20mA 출력
•하나의 활성 주파수 출력
HART
Bell 202 신호는 기본 mA 출력에 적재됨
장착 옵션
원격 장착 제어실만 해당
•35mm 레일에 장착
•레일은 접지되어야 함
적용 분야

•연속 공정 제어
•2단계 유량/혼입 가스
•충진 및 도징
•미터 검증
위험 승인
위험하지 않은 환경에만 적합

특징

Micro Motion H400S Hygienic Coriolis Meter, 4 Inch (DN100), 316L Stainless Steel, 32 RA Finish
  • Minimize variability in process control with superior sensitivity in a compact design
  • Ensure product consistency and quality with superior density measurement
  • Enable fast product change-over with self-draining design and superior temperature stability
  • Choose from options including highly polished 316L stainless steel with a surface finish up to 15 Ra
  • Gain real-time and in-process measurement integrity assurance with Smart Meter Verification
  • Realize superior two-phase flow capabilities for batching, loading and entrained air applications from compact, hygienic SST transmitter housings
  • Broad range of meter approvals and certifications, including; CSA (Class 1 Div 1 groups C and D, Class 1 Div 2 groups A, B, C and D), ATEX, IECEx, Ingress Protection 66/67, and NAMUR
  • Use in SIP, CIP and sanitary environments where 3-A compliance is required
Micro Motion 1500 Compact Control Room Single Variable Flow Transmitter
  • 35mm DIN 레일을 사용하여 분리 설치가 가능하여 배선 비용이 절감됩니다.
  • 제어실에 손쉽게 설치됩니다. - 0.03ft³ 공간만 필요(코리올리스 트랜스미터에 일반적으로 필요한 공간: 1.0ft³)
  • 전력 및 신호를 위해 센서에 간단하게 4-wire 연결로 구성이 간단합니다.
  • MVD 디지털 처리 기술을 통해 연행 가스 및 빈 관 내 배치 애플리케이션의 성능을 진단합니다
  • 정확한 현장 미터 검증 기능(in-situ meter verification)으로 공정 가동 중이나 보조 레퍼런스 없이 라인 내 센서의 상태 확인할 수 있습니다.

문서 및 도면

Micro Motion H400S Hygienic Coriolis Meter, 4 Inch (DN100), 316L Stainless Steel, 32 RA Finish
Micro Motion 1500 Compact Control Room Single Variable Flow Transmitter